GR2002 automatic high frequency heat sealing machine instruction manual

I. Overview
Liquid medicines such as blood and saline are gradually taking special plastic bags to fill, store and transport, and infusions replace traditional bottled methods. The sealing of the bagged liquid medicine requires that the sealing is clean, non-toxic, non-polluting, reliable, and convenient to operate. The heat sealing machine is designed for this purpose.
The design of this heat sealing machine is to listen to the opinions of the majority of users, improve the old heat sealing machine, and successfully develop it by the technicians carefully designing and repeating the test. Therefore, this heat sealing machine has the following characteristics compared with the old heat sealing machine:
1. This machine uses a microcomputer to control the heat sealing time. The heat sealing time can be precisely adjusted between 0.1 seconds and 9.9 seconds, so that the heat sealing quality is guaranteed.
2. This machine is equipped with elastic auxiliary device on the heat sealing head to prevent or reduce the hemolysis caused during the heat sealing process.
3. This machine replaces the traditional switch device with photoelectric device. It does not need to press the switch every time it is heat-sealed. It only needs to place the plastic tube directly between the heat sealing heads, and the heat sealing will be carried out automatically, which makes the operation simpler and more comfortable.
4. This machine can automatically adjust the grid voltage. When the grid voltage changes between 190V and 250V, it can work normally and smoothly. This shows the superiority of the machine when it is used in areas where power fluctuations are large.
Second, the main technical indicators
1, the power supply voltage: 190V ~ 250V, 50Hz
2, power supply: 200W
3, heat sealing time: 0.1 ~ 9.9 seconds adjustable
4, applicable pipe diameter: Φ3 ~ Φ6
5, size: 270 * 170 * 220mm
Third, use
1. Plug in the power supply and turn the power switch on the lower right corner of the instrument to turn on the instrument.
2. After the instrument is powered on, please preheat for 3 to 5 minutes and then heat it. If heat is applied immediately after energization, it will be detrimental to the service life of the instrument.
3. When heat sealing, the pipe is directly placed between the heat sealing heads, and the heat sealing starts. After the heat sealing is completed, the heat sealing head automatically jumps off, and the next working cycle can be performed later.
Fourth, maintenance and repair
1. When the heat is closed, the pipeline is in a loose state. Do not pull or shake the pipeline tightly, otherwise it will easily cause the pipeline to rupture and ooze. The instrument does not want to break the pipe in the heat-sealed state.
2. If there is bleeding or seepage during the heat sealing process, it should be resealed. In the off state, the stains are removed and dried in time to prevent the heat sealing head from being stuck and damaging the photoelectric sensing device.
3, automatic heat sealing in the test machine should avoid empty (no pipe) heat sealing operation, to prevent load short circuit, fire, so as not to burn the tube.
4, the ambient temperature is high and low, the pipe material is soft and hard, the need for the heat output of the heat sealing head is different. Time adjustment; for a long time, the power output lasts longer and the heat seal is a little more transparent. In short, the best heat sealing can be achieved by adjusting the working state appropriately for different situations.
5. The automatic heat sealing head has been carefully assembled and debugged before leaving the factory, and it is in the best working condition. Users should not disassemble it during maintenance and repair, so as to avoid unnecessary trouble.
Five, common fault phenomena and their elimination methods
Fault phenomenon
cause
Method of exclusion
No power to turn on the power
display
No city electricity
Check the mains
Fuse damage
Check and replace the 3A fuse
No heat
In-machine high voltage fuse is damaged
Check and replace the 0.25A fuse
Damage to the high pressure pipe inside the machine
Check and replace high pressure pipe
Heat sealing head without stroke
Photoelectron sensor port blocked
Check, alcohol wipes off dirt
The heat sealing head cannot be reset and the reset is not smooth
Miscellaneous or bloody dirt stuck in the activity gap
Shut down and handle debris and blood pollution in time
Other abnormalities in the heat sealing process
External interference
After shutting down for 7 seconds, turn it on again.